How to run an electrical safety test on an MRI in 10 easy steps with the ESA615 using the Point to Point function
Step 1
Plug the ESA615 into an electrical outlet
Step 2
Connect the alligator clips to both lead cords
Step 3
Plug the lead cord accessory (50’ or 75’) into the red input plug on the front of the ESA615
Step 4
Plug the lead cord with the black input jack into the black input plug on the front of the ESA615
Step 5
Put the alligator clips together and zero out the leads on the ESA615 by choosing Point to Point resistance.
Step 6
Insert the ground adapter into the electrical outlet for the ground reference point
Step 7
Connect the black alligator clip/lead cord to the ground adapter
Step 8
Run the long lead cord with the red alligator clip into the MRI room and clamp onto any exposed metal (ground point) on the MRI machine
Step 9
Run Point to Point to test function (volts, ohms, micro amps)
Step 10
Record results
You will need the following to perform this test:
Models/descriptions
4132046 ESA615 Electrical Safety Analyzer
(US/120V - alligator clamps included)
2392639 Ground adapter
4316223 50’ Lead cord
4794343 75’ Lead cord