NOTE: This version requires Ansur Executive v3.1.1
or newer.
Introduction
This read me file outlines
changes to each subsequent version of Ansur ESA615 Safety Analyzer relative
to the previous released version.
History
This chapter describes the
changes made between each new released version.
Release 1.1.2
- Units
modification to interpret L value as ‘mA’.
- Dialog
box failure text is the file is downloaded with a comma character in the
file name.
- Fix
for the “Live to Earth” and “Neutral to Earth” mismatch.
Release 1.1.1
- Support
for the Ansur Executive v3.1.1
- ZigBee
Communication update.
- Fix
for the error in uploading the test sequence with comma in the file
name.
Release 1.0.5
- Changed
STD, MAP commands for Mains on Applied Parts test element based on
standard selection.
- Sending
<LISO> tag for Mains on Applied parts test element, when AAMI
standard is selected in Mini Plug-In.
- Displaying
last modified print title in the Test Results of the ESA615 Mini
Plug-In.
- Added
MAP=3.5MA command for Direct Applied Part Leakage test element.
- Changed
MAP, MAP=7.5MA/MAP=1MA command for Mains on Applied Parts test element
based on lead standard selection.
Release 1.0.4
- Bug
fix for the resistance tests which displayed incorrect values
instead of XX or OL.
Release 1.0.3.1
- Changed
Italian language strings for Mini Plug-In.
Release 1.0.3
NOTE: This version
requires ESA615 Firmware v2.05 or newer
- Implemented
AP2 command for Patient Leakage Current and Mains on Applied Parts Tests to define
non-measured leads.
- Changed
the text Applied part “Class” to “Type”.
- Language
support for Italian and English languages.
- Retries
of REMOTE command if response is “!00” from
device.
- Bug
fix for adding “POWER-OFF” tag while uploading a template to Mini
Plug-In.
- Bug
fix for displaying unexpected error when the device stops responding.
Release 1.0.2
- Possible
to convert *.DTA files (from SD card) to either *.CSV or Ansur format
- Ansur
template are sent to the device in optimized form
- Sub
test elements of same test containers are combined while downloading
from the device
- Multiple
templates can be selected and send to device
- Enable
LOG serial data for the main and mini Plug-In
Release 1.0.1
NOTE: This version requires
ESA615 Firmware v2.02 or newer
- Removed
duplication of file names in the list view of the Mini Plug-In.
- Patient
Auxiliary Current test results got from device are updated with the
highest measurement values
- The
Mini Plug-In always overrides the Ansur template/ test sequence standard
to CUSTOM standard when sent to device, however differentiates between
IEC and AAMI
- Bug
fix to a problem of downloading blank CSV files and downloading files
when there are greater files on device, by introducing 3 seconds time
out
- Added
Multiple PE test functionality for Protective Earth
Resistance test while sending/downloading templates and test results
to/from the instrument.
- Restricting
file names to 50 characters while sending to the instrument
Release 1.0.0
- Initial
version of the ESA615 Plug-In.
Language Changes
This chapter describes the changes made to the language
file in each new version.
The list includes only versions where the language file was changed.
Release 1.0.0
- Initial
revision
Release 1.0.3
- Included
Italian language
Fluke Biomedical
6045 Cochran
Road
Cleveland,
OH 44139-3303 U.S.A.
Fluke Biomedical Europe
Science
Park Eindhoven 5110
5692EC
Son, The Netherlands
|
For more information, contact us:
U.S.A. (800) 850-4608 or
Fax (440)
349-2307
Europe/M-East/Africa
+31 40 267 5435 or
Fax +31
40 267 5436
Other
countries +1 (440) 248-9300 or
Fax +1
(440) 349-2307
Email: [email protected]
Web
access: www.flukebiomedical.com
© 2014
Fluke Corporation. All rights reserved.
|
|