This read me file outlines changes to each subsequent version of Ansur ESA620 Electrical Safety Analyzer relative to the
previous released version.
History
This section describes the changes made between each new released version.
Release 1.2.6
1. Fix
for the AP commands of Accessible
Leakage, Earth Leakage and Enclosure Leakage tests.
2. Fix
for the AP2 commands of Differential Leakage,
Earth Leakage and Enclosure Leakage tests.
Release 1.2.5
1. Fix
for the unexpected rendering of the polarity switch delay dropdown in the UI.
2. Corrected
the limits and changed the part type from ‘Body’ to ‘Body Float’ for
60601-2-4 templates.
Release 1.2.4
1. Notification
after equipment outlet change displayed with incorrect timing
Release 1.2.3
1. Fix
for sending AP / AP2 command based on the PCA_TYPE? response.
2. Retry
option for Calibration fail during a PE test has been fixed
3. UI
changes for polarity switch delay drop down
4. Retained
value for power on delay text box
Release 1.2.2
1. User pop up displayed for
tests where plug-in was changing the polarity without notifying the user
2. AP command modification.
Added //GND to the earthed tests.
3. Created 2 new templates
61010-CL1 and 61010-CL2
4. Transferred the limits from
60601 standard to the 60601-2-4 standard
5. Polarity switching delay
modified to be a non-zero value
6. Multiple PE tests returns
!57 and 10 seconds wait time is given for the transformer to cool down and
command is resent
Release 1.2.1
- Support for the
new Ansur Executive v3.1.1.
Release 1.1.14
NOTE: This version
requires ESA620 Firmware v2.10 or newer.
- Bug fix: Polarity
Switching Delay
Release 1.1.12
NOTE: This version
requires ESA620 Firmware v2.10 or newer.
- Bug fix: Display “OFF
to ON” in power configuration message if the equipment outlet is
switched from OFF to ON.
Release 1.1.11
NOTE: This version
requires ESA620 Firmware v2.10 or newer.
- Modified
AP2 commands to define non-measured leads.
- Bug fix: Display of
MTI data of all instruments in the print report.
- Modified
AP2 commands for IEC62353 tests.
Release 1.1.10
NOTE: This version
requires ESA620 Firmware v2.10 or newer.
- Implemented
new AP2 command to define non-measured leads
- Modified
AS/NZ standard templates in the Ansur Test
Library
Release 1.1.9
- Bug fix:
Stepwise start now works properly with the display of PLC settings
configuration on procedure tab.
- Bug fix:
Protect Earth Resistance test now shows proper results instead of OL
along with the test status.
- Bug fix:
Blank error message box is now not displayed for Protect Earth
Resistance when ‘Test Each Applied Part Separately’ is checked.
- Bug fix:
The Print report now not freezes when limit checkbox is un-checked and
given print for the first time.
Release 1.1.8.1
1.
IEC61010 template has been updated with
“Test status container” test.
Release 1.1.8
- Bug fix for the
Protective earth resistance test when it placed under the Test
container.
- Bug fix for Insulation
resistance test not displaying the standards information in the print
report.
- Bug fix for Insulation
resistance error during the second test.
Release 1.1.7.1
- Included User manual
Rev4.
Release 1.1.7
- Bug fix for B-Type
Applied parts (According to the standard IEC60601-1Ed3) for Patient
Leakage Current.
Release 1.1.6
- PLC and MAP tests –
“All leads” option is applicable only to CF lead type.
- MAP, PAC & PLC
tests – “Both” option should be enabled when worst case is selected.
- Bug fix for PAC test,
which passes even if Hi and Lo limit of the applied parts are crossed.
Release 1.1.5
- Implemented the “both”
option in Applied Parts of Patient Leakage Current, Mains on Applied
Parts & Patient Auxiliary Current.
- Implemented “Polarity
Switching Delay” option in preferences screen & Auto sequence.
- Implemented
IEC60601-1Ed3 for Patient Leakage Current, Mains on Applied Parts &
Patient Auxiliary Current.
Release 1.1.4
- Removed delays and
inserted 3 retry for the error responses.
- Used only Applied
parts Open for Patient auxiliary current test.
- Removed individual
applied parts measurements in the Insulation Resistance test.
Release 1.1.3
- Bug Fix for the
implementation of BF modules at 500V.
Release 1.1.2
- Implementation for
Insulation Resistance B Class modules and 500V – where "Applied Parts to Protective
Earth" and "Applied Parts to Non-Earth Accessible Conducive
Part " tests are not applicable.
Release 1.1.1.2
- Included the user
manual to the installation set.
Release 1.1.1.1
- Created installation
set with the latest DLLs avoiding DLL compatibility issues.
Release 1.1.1
- Removed “All Leads”
option for insulation Resistance Test.
Release 1.1.0
- Corrected the
alignment text for Italian Language.
- Implemented the “worst
Case” option for Patient Auxiliary Current.
- Implemented selection
for Module setup for the Sequence files and Job order files.
Release 1.0.8
- Patient Auxiliary
Current test implemented to measure more than or equal to two leads.
- Bug fixed for Test
lead Calibration.
- Updated IEC templates.
- For MAP voltage, text
changed from 25A to High and 200mA to Low.
- Implemented to
overcome error !37.
Release 1.0.7.2
- Added the user manual
and the Italian language file to the installation set.
- Removed the Spanish
templates from the installation set.
Release 1.0.7.1
- Added the Spanish
templates in the installation set.
Release 1.0.7
- Power OFF delay dialog
appears when complete test is finished.
- The test result status
will update only when all the sub tests under the Test Group are
finished.
- Configuring Patient Lead
configuration in the custom setup.
- Option for Power OFF
delay message, before or after switching Power OFF.
- Multiple module
configurations can be created in a template.
- Bug fixed for
simulation of Differential Leakage test and ECG test.
- MAP polarity and MAP
voltage levels can be selected in the Custom Setup.
Release 1.0.6
- Bug fixed for the
module names displayed in VISTA.
- Templates are updated
for Class II in IEC standards.
- Bug fixed for the
results display of CF modules.
Release 1.0.5
- Implemented the "Patient
Lead Configuration" can configured in the custom setup.
- Added IEC 61010 and
IEC 60601-2-4 templates to the template library.
- Test guide displays
all measurements for multi-module test.
- MAP voltage levels and
polarity can be selected in the custom setup.
- Insulation resistance
test duration can be selected in the custom setup.
Release 1.0.4
NOTE: This version
requires ESA620 Firmware v1.17 or newer.
- Implemented additional
remote control command for correcting the MAP and Patient Leakage
Current measurement.
- Corrected Alternative
Equipment Leakage Current measurement for Open Earth
Release 1.0.3
- Corrected errors in
performing the test lead calibration for 200mA Earth Resistance test.
- Corrected inaccuracies
in test limits for the sample test templates.
Release 1.0.2
- Improved
communications timing during the 25A Earth Resistance test.
Release 1.0.1
- Corrected a problem
with zeroing leads for resistance measurements when the number of leads
was different than the default set on the instrument.
- Increased communication
timeout.
- Modifications to
French language file.
Release 1.0.0
- V1.0.0 is the first
release of the ESA620 Plug-In.
Language
Changes
This chapter describes the changes made to the language file in each new
version.
The list includes only versions where the language file was changed.
Release 1.1.5
- Added Language IDs:
20188, 20264 to 20268, 20794 to 20798, 21102.
- Modified language IDs:
20256, 20257, 20259, 20261, 20306, 20342, 20500, 20501, 20503, 20510 to 20515,
20524, 20525, 20530, 20533, 20534, 20537, 20701, 20731 to 20733, 20767
to 20769, 20786, 20789, 20790, 20792, 20793
Release 1.1.0
- Added French and
Italian Language IDs: 20206 to 20207, 20770 to 20793, 20826 to 20835 and
21022.
Release 1.0.7
- Added: 20206 to 20207,
20770 to 20793, 20826 to 20835 and 21022.
Release 1.0.5
- Added: 21909
Release 1.0.4
- Added: 20206, 20770 to
20790, 20826 to 20835
Release 1.0.1
- Added: 20161
Release 1.0.0
- Initial revision
|